Count Rate Performance Ratio Trues/Randoms unbalanced for high Activities Processing of detected photons takes finite time Noise Equivalent Count Rate: fR ST T NECR 2 2 T : trues
S : scattered R : randoms f : rel. area of object on proj. surface
Count rate curves for true, random and multiple coincidences. Derived curves for expected and noise equivalent count rate [2]. Data recorded on a CTI ECAT 953B PET using a 11 C-filled cylinder in water.
Maximum image quality
Low image contrast
Noisy image
26
ESTRO Teaching Course Advanced Imaging: PET – Basic Principles