ASMR 2016
Set-up accuracy: interfraction motion
Positioning system Imaging modality Position error 2D-2D image registration for verification of set-up based on bony anatomy registration
Study
2.3mm ± 1.6mm
Rosenthal 1995
Dental fixation
Ortogonal radiographs
Vogely Bale Hohner head Holder
1.9mm ± 1.2mm 1.8mm ± 0.8mm 1.3mm ± 0.9mm
Sweeny 2001 Kumar 2005 Georg 2006
Portal imaging Portal imaging Portal imaging
Gill-Thomas-Cosman
!
Brain Lab Mask
3D-3D image registration for verification of set-up
3.7mm ± 0.8mm
Baumert 2005
Stereotactic mast
CT
Boda-Heggermann 2006 Guckenberger 2007
3.1mm ± 1.5mm 3.0mm ± 1.7mm 2.9mm ± 1.3mm 3.2mm ± 1.5mm
Scotch cast mask Scotch cast mask
CBCT CBCT CBCT CBCT
Thermoplastic mask & Bite block
Masi 2008
Bite-block
Courtesy M. Guckenberger, ESTRO IGRT course