Independent MU checks
• Recalculates the dose, based on the plan parameters from the planning system (or v.v.)
• This will check (in theory) – Amount of monitor units
– Problems with plan normalization – Computation errors of planning system
• Third party software
– Lots of software around (small companies) – Check what it really checks – Test with intentional errors