ASMR

Process map: Online 3D portal imaging

e.g. 6) Image acquisition: • Image field treated, PI software wasn’t ready yet • Additional dose of 3 cGy put into patient • Likelyhood: 3-4 • Severity: 2-4 • Detectability: 3

• RPN=3-4*2-4*3=18-48 • Recommendations:  Education  Better software

Made with