Type B uncertainty … input Source geometry Information is required on geometry, materials, density, elemental composition, and dimensions. The uncertainty of this information is more crucial for low energy sources For reference dosimetry of new sources geometry and dimensions should be verified experimentally in a sample of sources. If the source includes parts of non-negligible mobility, MC should be performed for different configurations and results averaged.
contact tr. radiography (IsoSeed I25.S17plus)
seed cut out view (3631 A/M)
e microscopy (6711)
Pantelis et al, J Contemp Brachytherapy 5(4), 240 (2013)
Pics from: Rogers & Cygler (Eds), Clinical dosimetry measurements in radiotherapy (2009 AAPM Summer School), Monograph No. 34, Medical Physics Publishing 2011