Advanced Physics for Brachytherapy 2018
Accuracy of source localisation
CT phantom (solid) Siebert et al. R&O 2007
reconstruction uncertainty (1 SD) < 1.4 mm for 4-5 mm scans < 1.0 mm for 2-3 mm scans
MRI / CT phantom (agarose gel) De Brabandere et al. R&O 2006
uncertainties for MRI slightly larger than for CT reconstruction uncertainty < 2 mm for 3-5 mm scans
See also DeBrabandere et al. 2011,
Made with FlippingBook - professional solution for displaying marketing and sales documents online