Seed/Applicator Model Accuracy Requirements
• Patient CT grids (>1 mm voxel) are probably not adequate for accurate modeling on the spatial scale of brachytherapy sources and applicators.
• MBDCA vendors should use analytic modeling schemes or recursively specify meshes with 1–10 μ m spatial resolution.
• Vendors to disclose their geometry, material assignments, and manufacturing tolerances to both end users and TPS vendors (if responsible for data entry and maintenance)
TG-186 Section IV-B