Advanced Physics for Brachytherapy 2018
Beam quality:Q 0 (Co-60 or MV linac photon beam)
Calibration: !!! Calibration uncertainty will be propagated as a type B unc. component to experimental results !!!
Absorbed dose sensitivity calibration
) 0( QM
1
dosimeter
( , Q
)0 =
=
S w AD
D
)0 QwADN Q ( ,
)0(
w
is usually performed using a Linac where: M(Q 0 ) is the dosimeter reading
D
(Q 0 ) is dose to water at the point of measurement in the absence of the detector obtained using an established reference dosimetry protocol (e.g. TRS398) and an ion chamber with a dose to water calibration traceable to international standards
w