Dose in ICRU83
Level 1. Why is it not adequate today?
-The absorbed dose distribution for IMRT can be less homogeneous then in CRT
-Each beam can produce absorbed dose with large dose gradients
- Large dose gradients (10%/mm) in the PTV boundary i.e. small shifts in delivery can affect the reliability of using a single point to report the dose
- Because modern TPS have evaluation tools that makes it possible.
- Monte Carlo calculations have statistical fluctuation in the results for small volumes which makes it difficult and uncertain to determine an absorbed dose to a point.