CPB Ljubljana 2018 course book
Dose calculation – AAPM-TG43: S K
and
The product of the actual source strength S K and the dose rate constant Λ of the source and radionuclide considered is the actual dose rate at 1 cm radial distance at θ = θ 0 = 90 0 .
0 )= (1cm, 90 0 )
(r
, θ
0
r
= 1cm
0
Λ is expressed in units of cGy.h -1 .U -1 and 1U = 1 µGy.m².h -1 = 1 cGy.cm².h -1 Dose rate is in cGy.h -1