Course book Advanced Treatment Planning 2018
Why switch to MC dose computation?
PB algorithm (especially in combination with range shifter) inaccurate for two reasons:
➢ Nuclear halo effect • Each pencil beam is modelled by 2 Gaussians (MCS, nuclear halo) • Lack of handling nuclear halo properly within the range shifter, then transporting the beam through vacuum (instead of air) and large heterogeneities (patient surface): causes lack of modelling accuracy especially for low energies where a greater angular spread of the protons is expected. ➢ Lateral heterogeneities • Each spot is split into 19 sub-pencil beams.
Courtesy N. Schreuder, Provision Knoxville, 2017
• In case of large spot sizes (combination of range shifter and larger gaps) the distance between subspots becomes larger than anatomic density variations within the patient.
Source: RS5 reference manual, RSL