Lessons learned from Epinal
Potential errors
o Wrong use of TPS due to lack of training + unsafe screen display #1 o Dose due to verification imaging (MV portal) not taken into account #2 o Calculation error due to in-house software, not tested, not qualified #3
Sole physicist
o
Prevention
o Time and organisation for continuous training
o Team of physicists (at least 2)
QA for software
o
o Software with safe human-computer interaction
o In vivo dosimetry and second independent calculation
Utrecht 2016