Dose 2016
Energy dependence due to changing Stopping-power ratios
6 MV photon beam n=narrow beam ( ∅ =0.5 cm) b=broad beam ( ∅ =10 cm)
Cylindrical air ion chambers can be used without correction for energy variations in high energy photon beams, incl. Co- 60 (IAEA TRS-398).
The secondary electron energy spectrum is not changing with depth.
Heydarian et al
Ratio narrow/large field depth dependence Difference <0.5%
Made with FlippingBook