Dose 2016
Organic scintillator (W1 Standard Imaging)
OF: small fields
Perpendicular to beam axis
Parallel to beam axis
5.0%
4.0%
3.0%
The calibration methodology proposed by the manufacturer is not appropriate for measuring output factors for small fields. Calibration using smaller fields and the detector axis parallel to beam axis should
2.0%
1.0%
Exradin W1 Cal. A Exradin W1 Cal. B Exradin W1 Cal. C
0.0%
-1.0%
be used.
Pinpoint
-2.0%
These results show that the CRL and gain depend on field size and orientation of the scintillator.
-3.0%
-4.0%
Relative difference normalized to EFD OF
-5.0%
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0
Side of equivalent square field size at isocenter (cm)
Relative differences of OF normalized to EFD-3G OF.
OF are referred to a 10 x 10 cm 2 field size.