Dose 2016
Detector
Appropiate for
Be careful
For x-ray depth dose measurements
Check polarity effect for both photon and electron beams
Ionization chambers
Plane-parallel chamber for build- up region
Always measure with the same nominal V
For micro-chambers the stem signal can be an issue Correct by Sw,air for electron depth dose measurements.
Output factors large fields
Output factors for small fields (micro-chambers)
Use small volume detectors for scanning
Effective point of measurement
Check stability temperature during measurements
Depth dose measurements and profiles
Energy dependence
Diodes
Sensitivity variation with accumulated dose May present sensitivity variation with dose rate
Shielded detector for x-rays
Non-shielded for electron beams
Output factors
Check stability temperature during measurements
Shielded detector standard/large fields
Non-shielded for small fields