Dose 2016
Origin of peripheral dose contributions Anders A .
Patient scatter
– Unavoidable – Decrease from target essentially as:
µ − ⋅
r
n =1 column shaped scatter source
e
n =2 “blob” shaped scatter source
n
r
Head scatter
− Sources are all irradiated parts of the treatment head, main source the flattening filter
Limited by the collimating devices
−
− Machine design dependent (FFF machines have less!)
Leakage, including scatter leakage
− Depends on collimator and shield thicknesses and material (density)
Collimator design dependent
−
Treatment technique dependent
−
Neutrons
− Mainly a high Z phenomena at photon energies above photonuclear threshold values
Avoidable by lower beam energy
−
Only one factor user controllable on a daily basis, all other once per decade (replacement)!