Tolerance limits and action levels
2 '
2 '
2
'
'
u
u
u
u + +
=
total
t measuremen
Fcal
CF
“The overall accuracy of entrance in vivo measurements is determined by the combined uncertainty in the calibration factor, the correction factors and the accuracy and reproducibility in diode position ” In our case it is equal to 2% (1sd ). So without any additional cause of error 68% of all measurements would have a dispersion of ± 2% and 95% of ± 4%.”
Essers and Mijnheer, Int.J.Radiat.Oncol.Biol.Phys., 1999