Calibration Process for DVS
Lot Dose Response - RADFET Radiation Sensitivity
• A cumulative dose response calibration curve is obtained for a specific lot by irradiating a statistically significant representative sample from the lot up to 80Gy ( maximum dose range of the dosimeter ).
0.5
0.45
0.4
0.35
0.3
(mV/cGy)
0.25
0.2 Radiation Sensitivity
201 1004 2008
3008 4009 5011 6018 7026 8036
Cumulat ive Dose ( cGy)
Dose response curve
Verified by UW-ADCL (sample lot sent for testing)
❑ Calibration is valid for use with daily doses of 150-250 cGy
❑ Reported accuracy for each lot has a calibration certificate with values within:
❑ <5.5% (2σ) up to 20 Gy
❑ <6.5% (2σ) up to 74 Gy (accuracy decreases for doses > 74 Gy).
Joanna Cygler, MOSFET dosimetry, AAPM Summer School, 2009