Errors in radiation oncology: A study in pathways and dosimetric impact Eric Klein, Robert E. Drzymala, James A. Purdy and Jeff Michalski Journal of applied Med. Phys. 2005
Error type:
Typical # of fraction in which it was detected
Method of detection
# of events
#of record and verify
Incorrect treatment coordinate(s) Wrong gantry angle Wrong or omitted cerrobend block Wrong field size Incorrect collimator angle Missing compensation filter Incorrect calculation
first
Port film
19
7
first first
Port film Port film
15 15
8
Errors over a 30 months on 7 linacs equiped with R&V system . No complete electronic transfer of treatment set-up data.
N/A
2 to 5 fractions
In vivo dosimetry; physicist chart review
11
N/A
first first
Port film. Port film
9 8
8 3
2 to 5 fractions
In vivo dosimetry; physicist chart review
6
N/A
Incorrect MU Wrong photon energy
2 to 5 fractions 2 to 5 fractions
In vivo dosimetry In vivo dosimetry
5 3
4 3
Missing or incorrect MLC shape
first
Port film
3
Detected errors 49/3900
Incorrect wedge direction
1,5,16 fractions
Later diode check; physics check Physicist review
3
Incorrect number of fractions for given set of fields Incorrect or rotated compensating filter Patient treated head to gantry but scanned foot to gantry
1 to 3 fractions
3
2
With entrance in vivo dose measurements 30/3900
2 or 11 fractions
Therapist discovery; in vivo dosimetry
2
N/A
First fraction
Port film
1
N/A