Causes of differences between TPS & measured dose Source Unintentional Errors
Accepted Limitations
Beam data incorrectly measured such as small field output factors, MLC Transmission
Accepted TPS Beam Model Limitations ex.T&G modelling inaccurate
TPS
Settings changed accidently ex. gantry resolution for VMAT calculation
Continuous treatment segmented into static fields at a pre-set gantry resolution
Dynamic Gantry Tolerance = 0.5°
Data transfer errors
Dynamic MU Tolerance = 0.2 MU
Dynamic MLC Tolerance = 5mm
Linac
Linac component fall’s out of specification (MLC calibration, Energy, Flatness/Symmetry)
Tolerance on Symmetry/Flatness with rotation and @ low DR’s
Tolerance on MLC calibration