Organic scintillator (W1 Standard Imaging)
• Good water equivalence for high energy photon and electron beams
Result
Uncertainty
Test
Short-term repeatability (@0.75 Gy) Short-term repeatability (@0.15 Gy)
σ = 0.10%
0.07%
σ = 0.25%
0.05%
Dose-response linearity
RMS = 0.61 %
0.20%
Angular dependence
RMS = 0.21 %
0.07%
-0.225 %·⁰C -1
0.008%·⁰C -1
Temperature dependence
Time to reach thermal equilibrium*
1 min 40 s
16 s
Repetition rate dependence
RMS = 0.53 %
0.06%
Deviation from ISL
RMS = 0.38 %
0.26%
-0.28%·kGy -1 [0-15 kGy] -0.032%·kGy -1 [15-127 kGy]
0.06%·kGy -1 0.018%·kGy -1
Loss of sensitivity with accumulated dose
Made with FlippingBook HTML5