Dose Course 2018_Flipping book
Detector
Appropiate for
Be careful
❑ For x-ray depth dose measurements
Check polarity effect for both photon and electron beams Always measure with the same nominal V For micro-chambers the stem signal can be an issue Correct by Sw,air for electron depth dose measurements. Use small volume detectors for scanning
Ionization chambers
❑
❑
Plane-parallel chamber for build-up region
❑
❑
❑ Output factors large fields
❑
❑ Output factors for small fields (micro-chambers)
❑
Effective point of measurement
❑
Check stability temperature during measurements
❑
❑ Depth dose measurements and profiles
Energy dependence
Diodes
❑
Sensitivity variation with accumulated dose
❑
❑
Shielded detector for x-rays
❑
Non-shielded for electron beams
❑ May present sensitivity variation with dose rate
Output factors
❑
Check stability temperature during measurements
❑
❑
Shielded detector standard/large fields
❑
Non-shielded for small fields