Factor-based dose calculations: basic dosimetric quantities
S c : mini-phantom/buildup cap design – depth considerations
The effective measurement depth in the mini-phantom/build-up cap must be large enough to stop contaminating electrons.
24 MV
6 MV
d=4 cm
d=5 cm
d=10 cm
d=0.9 cm
d=2.3 cm
Two rules of thumb used to eliminate e - contamination:
d 2· d max d MV/3 [cm]
Frye DM, et al (1995) Med Phys 22, 249-53