Dose per MU formalisms: factor-based dose calculations
Modulation with physical wedges
1,15
Example: Elekta Precise, 60 ° internal wedge
1,10
X 6 - op
Variation open vs wedged fields:
X 10 - op
With FS
X 25 - op
1,05
➔ ~ 10% open
X 6 - we
➔ ~ 20% wedge
X 10 - we
1,00
X 25 - we
Wedge angle
Energy
0,95
OUTPUT RATIO IN AIR
Linac specific
0,90
0 5 10 15 20 25 30 35 40
EQUIV. FS [cm]
• Physical wedges introduce changes in the beam spectrum, which are dependent on wedge material and influence dosimetric parameters that vary with depth (TPR, RDD) as well as phantom scatter, Sp • The position of the wedge, whether internal (motorised) or external (manually inserted) affects Sc
• Dosimetric parameters for wedged beams should not be confused with open beam data • Irregular wedged beams need some special considerations for MU calculation / verification
Additional correction factors needed in the dose per MU formalism