Dose Course 2018_Flipping book
Radiochromic film evolution…
POST EBT GAFCHROMIC FILMS (2004) [EBT;EBT2;EBT3]
PRIOR EBT GAFCHROMIC FILMS [HD-810; MD-55-2; HS]
❑
Low sensitivity
Dose range 2cGy – 8 Gy
❑
❑
Small size
Saturation ~ 10Gy (15Gy)
❑
❑ Inherent optical density non-uniformity in one sheet (15%) [Need of double-exposure]
Different sizes available
❑
Z eff ~ 7.05 (chlorine addition)
❑
Significant better homogeneity
❑
❑
High price
[no need of double-exposure techniques]
Lower price
❑