Dose Course 2018_Flipping book
Calibration: Determination of the sensitometric curve
❑ The film calibration procedure consists in acquiring a set of single measurements of net
optical density NOD with the corresponding absorbed dose and obtaining a function
which relates NOD for a given dose, called the sensitometric curve. The NOD is obtained
by subtracting the average OD of a non-irradiated film to an irradiated film using
regions of interest ROIs of constant sizes and uniform field doses
OD =-log
(T) =-log
(I
/I)
10
10
0
Made with FlippingBook HTML5