1. Basic dose algorithm validation TPS model comparison tests

o Needs to be completed for each configured beam.

o Each physical wedge is a unique beam (different energy fluence spectrum) o Nonphysical wedges can be considered an extension of the correspondent open field (is not a different beam)

o No extra measurements are needed. Confirmation that the dose calculated in the planning module agrees with those in the modeling module.