Dose Course 2018_Flipping book
Collimator leakage
MLC intraleaf and interleaf
1. Two effects: • Diffused dose from spiky interleaf fluence leakage • Intraleaf attenuation
Intraleaf leakage very small:
m −
t
0.0408 18.0 8.0 −
e
→
e
=
0.28%
8 cm tungsten at 3 MeV
Measuements from Arnfield et al , 2000 Med. Phys. 27 p 2231-2241