20 6MV plans for different sites (breast, head and neck, pelvis, brain) were selected. Phantom dose distributions were calculated on Monaco 5.51 TPS using Monte Carlo dose calculation algorithm, irradiated on Elekta Versa HD linac and measured with both MatriXX devices in the appropriate phantoms, positioned on the isocenter positioning marks on them for measurement (on the linac couch) and in the tps (for plan isocenter). For evaluation purposes isocenter dose planes were exported from TPS as well as planes with vertical shift 1, -1, -2 and -3 mm. Gamma analysis 3%, 3mm, 5% threshold was performed for all measurements against all TPS planes for each 2D array. To evaluate the shift between the effective plane of measurement and setup plane for beams coming laterally, measurements of narrow 25x1 beam at 90 degrees gantry and 7.5 degrees collimator tilt were performed with both Evolution and Resolution, positioned with marks in the isocenter and vertically offset. Results Best mean gamma pass rate for Evolution was achieved with the shifted 2mm down TPS planes (mean 99.33%, min 96.2%, max 100%) and for the setup plane (no shift) for Resolution (mean 99.48%, min 97.0%, max 100%). In g90 tilted narrow beam measurements with Evolution: offset in Y-axis peak at isocenter, symmetrically offset on the Y- axis peak in 4 mm up (c); symmetrical around the center peak in 2 mm up (d) - effective plane of measurement and setup plane are not coinciding for lateral beams. Resolution did not exhibit the same behaviour: a symmetrical around Y peak with position in isocenter (b); almost symmetrically offset peaks at 2 mm up and 2 mm down (a).

Conclusion Effective plane correction is required for VMAT QA with MatriXX Evolution and can be achieved by using for verification the TPS plane 2 mm down from the setup plane, while no such correction is required for Resolution. Evolution, although introducing the gantry angle sensor and angular correction factors lookup tables in the software to compensate for the angular dependency, still uses the same basic design geometry and detectors as its predecessors, designed for measurement of beams perpendicular to the detector array, while the new Resolution has smaller detectors (smaller depth correction) and its geometry has been redesigned with rotational plan QA in mind, which could explain these results.

PO-1578 WIDMApp, an innovative approach for individual dose monitoring in Molecular Radiotherapy

C. Mancini Terracciano 1 , F. Collamati 2 , R. Faccini 1 , G. Iaccarino 3 , R. Mirabelli 1 , F. Nicolanti 1 , M. Pacilio 4 , A. Soriani 5 , L. Campana 6 , E. Solfaroli Camillocci 7 1 Sapienza, University of Rome, Physics, Rome, Italy; 2 INFN, Roma, Rome, Italy; 3 IRCCS Regina Elena National Cancer Institute, Laboratory of Medical Physics and Expert Systems, Rome, Italy; 4 Azienda Ospedaliera-Universitaria Policlinico Umberto I, Medical Physics Division, Rome, Italy; 5 IRCCS Regina Elena National Cancer Institute, Laboratory of Medical Physics and Expert System, Rome, Italy; 6 Sapienza University of Rome, Specialty School of Medical Physics, Rome, Italy; 7 Sapienza, University of Rome, Specialty School of Medical Physics, Rome, Italy Purpose or Objective We will present the proof-of-concept of WIDMApp (Wearable Individual Dose Monitoring Apparatus), a multi-channel detector and data processing system, aimed at allowing an accurate determination of absorbed dose per organ in Molecular Radiotherapy (MRT). The precise estimation of dose absorbed by the target and healthy organs is essential to increase therapeutic effectiveness, optimize the treatment planning and establish dose-effect relationships for tumor and normal tissues.

Materials and Methods

Made with FlippingBook Digital Publishing Software