ESTRO 2023 - Abstract Book

S1795

Digital Posters

ESTRO 2023

Using a CIRS IMRT thorax phantom, 3 spherical targets of volume 5, 18 and 40cm ³ were planned in Pinnacle ³ v16.4.3 to 54Gy in 3# in accordance with UK SABR Consortium guidelines [3]. Planning method was based on a local VMAT class solution using the medium setting to produce simple, open segments. Optimisation was varied using low, medium, high and very high complexity settings to produce four base plans per PTV. Plans were copied and adjusted by closing both MLC banks up to 1mm in 0.1mm increments, then doses recalculated at a 1mm resolution for a total of 132 plans. The planning study was validated experimentally for the 18cm ³ target, measuring all plans on a Truebeam using a MatriXX chamber array (iBA), reconstructed using Compass (iBA). Compass uses an independent beam model and TPS-class collapsed cone algorithm to reconstruct delivered dose distribution based on MatriXX measurements. Results MLC aperture reduction had a linear effect on target DVH statistics; however, the rate of degradation per mm closure depended on the target and optimisation setting used. In the worst case, the 5cm ³ target planned using the v. high setting resulted in 10% reduction in PTV D95% per mm closure per MLC bank. Higher complexity settings led to greater degradation under error. They also resulted in smaller mean field and lower modified gradient index (MGI) [3]; a key indicator of SABR plan quality.

Compass results for degraded plans relative to the original plans showed good agreement within 1% of TPS predictions.

Conclusion Optimisation parameters affect the robustness of SABR plans to machine delivery errors. Planners should carefully consider plan optimisation settings for SABR, especially when targets are small (<20cm3). Machine delivery errors below action tolerances may occur resulting in suboptimal plan dosimetry, and planners should consider this alongside the goal to reduce MGI. [1] Nithiyanatham, K. et al. (2015) “Analysis of direct clinical consequences of MLC positional errors in volumetric- modulated arc therapy using 3D dosimetry system”, Journal of Applied Clinical Medical Physics, Vol. 16, No.5, p.296-305 [2] Norvill, C. and Jenetsky, G. (2016) “Clinical significance of multi-leaf collimator calibration errors”, Australasian Physical and Engineering Sciences in Medicine, Vol 39, p. 253-258

[3] UK SABR Consortium (2019) “Stereotactic Ablative Body Radiation Therapy (SABR): A Resource”, Version 6.1

Made with FlippingBook - professional solution for displaying marketing and sales documents online