ESTRO 2023 - Abstract Book
S1537
Digital Posters
ESTRO 2023
1 H. U. La Princesa, Medical Physics, Madrid, Spain; 2 H. U. La Princesa, Radiation Oncology, Madrid, Spain
Purpose or Objective Dose calculation algorithms allow the estimation of the absorbed dose by the patient during radiotherapy treatment. Any changes to these algorithms should be characterised before they are used in the clinical process. The aim of this work is to show the dosimetric discrepancies observed in Anisotropic Analytical Algorithm (AAA), Acuros External Beam (AX) and Electron Monte Carlo (MC) calculation algorithms after a change in the secondary source modelling carried out in Eclipse version 15.6.6. Materials and Methods The work has been carried out in the process of software change between Eclipse versions 15.6.5 and 16.1.2, analysing the validation tests of calculation algorithms and the periodic quality control of the TPS. For the validation of the calculation algorithms, the simulated dose was recorded at a reference point, in irregular and asymmetric fields and fields with variations of SSD in photons and, varying the applicator and the SSD in electrons. For both photons and electrons, the procedure has been performed for all available energies. The treatments have been simulated in a Test-Box using a phantom with a CT value of -4 HU and water as the assigned material. In the periodic quality control of TPS, the constancy of the maximum, minimum, mean and median dose obtained after the simulation of a plan of 25 fractions of 2 Gy is analysed, using a calculation of version 15.6.5 as a reference. The phantom used is the "QUASAR Multi-Purpose Body Phantom" with 2 cedar wood inserts on the sides and the field configuration consists of 2 opposite lateral static fields, performing the simulation with all available energies. Results In the validation of the calculation algorithms, for the plans simulated with AAA, the deviation exceeds 0.5% between the versions analysed in irregular fields, asymmetric fields and fields with SSD variation, reaching a maximum deviation of 1.7% for asymmetric fields, as shown in image I. For the simulations with AX we only obtained deviations greater than 0.5% in asymmetric fields, with a maximum deviation of 1.5%. For the simulations with MC the deviations are larger than for AAA or AX, obtaining a maximum deviation of 2.02%.
In the periodic quality control the maximum discrepancies between versions are 1.5 Gy in one of the cavities for the minimum dose with AAA, 0.2 Gy for the maximum dose with AX in the same cavity and 1.6 Gy for MC in the maximum dose in the other cavity, as shown in table I.
Made with FlippingBook flipbook maker