PD-0982 Dosimetric and clinical aspects of lung SABR delivered with Synchrony® tracking on Tomotherapy M. Cerrato 1 , E. Gallio 2 , F. Menegatti 1 , I. Bonavero 1 , C. Grossi 1 , C. Casale 1 , P. Brossa 3 , S. Badellino 1 , F.R. Giglioli 2 , R. Parise 1 , M. Levis 1 , U. Ricardi 1 1 University of Turin, Department of Oncology, Turin, Italy; 2 A.O.U. Città della Salute e della Scienza, Medical Physics Unit, Turin, Italy; 3 A.O.U. Città della Salute e della Scienza, RTT Member, Turin, Italy Purpose or Objective In 2020 Synchrony® was installed on Radixact® tomotherapy at our department. This study compares the Radixact® plans and the LINAC-based lung SABR plans to evaluate possible SABR-induced toxicities resulting from the two different techniques. Materials and Methods Patients with early-stage NSCLC or oligometastatic lung tumors, PS ECOG 0-1, with a single nodule, in peripheral location were included. All patients completed the simulation phase successfully and were treated with the Synchrony® plan conforming to a “risk adapted” fractionation schedule (3-5 fractions). A LINAC-based plan was prepared for back-up. A dosimetric comparison was made between Radixact® tomotherapy and LINAC-based plans (Elekta® Versa HD). Beam-on time (BOT), PTV parameters (volume, volume at the prescribed dose, mean dose, and conformity number index), EQD2Gy mean dose (Ipsilateral lung – PTV), and volume (%) at different dose levels for (Lungs-PTV) and (Ipsilateral lung – PTV) (V5Gy, V10Gy, V13Gy, V15Gy, V20Gy and V25Gy) were considered. The volume of the isodose of 10 Gy, 15 Gy, 18 Gy, 20 Gy, and 30 Gy were also compared. Radiological toxicity was evaluated according to the Ronden et al classification of lung SABR outcomes. A Wilcoxon signed-rank test was applied for all parameters and a level of significance of 5% was considered. Results From July 2020 to July 2022, 18 patients were enrolled. The median beam-on time (BOT) for each fraction was 9.3 minutes (IQ range: 8.6-10.5). The following dose-volume parameters were statistically different between the two plans: BOT, (p<0.001), PTV volume (p<0.001), conformity number index (p=0.04), and volume of 15 Gy (p=0.002), 18 Gy (p<0.001), 20 (p<0.001) and 30 Gy (p=0.02) isodoses (Figure 1). With a median follow up of 15 months, approximately 67% of patients displayed radiological toxicity, mostly asymptomatic or with mild symptoms: 5 diffuse consolidations, 4 diffuse Ground Glass Opacities (GGOs), 3 patchy GGOs, 16 late scar-like fibrosis, 1 late mass like-fibrosis. These radiological outcomes have different features and a faster onset than in patients treated with LINAC-based plans, which have been employed extensively at our institute.

Conclusion LINAC-based lung SABR gives rise to a variety of toxicities that are strongly dependent on both patient-related factors and planning/dosimetry-related factors. There are no studies investigating the toxicity of lung SABR with Synchrony® tracking on Radixact® Tomotherapy. Preliminary data highlight a statistically significant difference in terms of volumes of isodose and suggest a possible correlation between these volume parameters and the development of SABR-related radiological toxicity. No differences in terms of clinical toxicity were found between patients treated with the two techniques. Further studies including a larger number of patients and longer follow-ups are necessary to draw more robust conclusions. PD-0983 Mid-ventilation vs ITV for SABR on the MR-Linac for locally advanced pancreatic cancer J. Gough 1,2 , R.A. Mitchell 3 , A. Dunlop 3 , D. McQuaid 1 , S. Nill 3 , S. Alexander 1,2 , H. McNair 1,2 , U. Oeflke 3 , B. Ng-Cheng-Hin 1 , A. Hunt 1 , K. Aitken 1,2 1 Royal Marsden Hospital, Radiotherapy, Sutton, United Kingdom; 2 Institute of Cancer Research, Radiotherapy and Imaging, London, United Kingdom; 3 The Royal Marsden Hospital and the Institute of Cancer Research, Joint Department of Physics, London, United Kingdom Purpose or Objective For Stereotactic Ablative Body Radiotherapy (SABR), it is desirable to limit target volumes to deliver ablative doses whilst sparing surrounding organs at risk (OARs). Internal target volumes (ITVs) may be used to account for respiratory motion if advanced motion management methods are not available. Delivering escalated doses may improve survival for locally advanced pancreatic cancer (LAPC), but safe SABR delivery requires limiting dose to adjacent GI OARs. Mid-ventilation (MidV) PTV margins have been shown to be dosimetrically advantageous for delivering conventionally fractionated RT (CFRT)to LAPC(1). This planning study compared ITV with MidV PTV margins for MR-Linac (MRL) SABR delivery for LAPC.

Made with FlippingBook flipbook maker