ESTRO 2024 - Abstract Book
S3176
Physics - Detectors, dose measurement and phantoms
ESTRO 2024
Figure 2 Dose values measured and calculated using the vertebrae phantom for different radiation treatment plans in relation to the reference values measured with the scintillator. mv = measured value, rv = reference value.
Conclusion:
The applied 3D printing allowed the fabrication of a vertebral phantom with density inhomogeneities of a bone-like structure. The feasibility of QA measurements in this phantom was demonstrated. The scintillator is recommended as the detector of choice for plan-specific QA in such a geometry. The unshielded diode and microDiamond detector proved effective in open fields but showed limitations in offset fields. Consequently, it can be inferred that they are less suitable or that the measurement results must be corrected to a greater extent.
Keywords: dose validation, detectors