Figure 2 – Dose difference between ion chamber and EPID corrected (with specific and global fit) and uncorrected reading

Conclusion:

A stable global reading-to-dose conversion is determined for aS1200, dependent on the field size. Accuracy in the dose determination is similar in the cases of global or detector-specific conversion factors. However, a missing correction would lead to dosimetric errors unacceptable if the detector is used for dosimetric purposes.

Keywords: EPID, reading-to-dose conversion

References:

[1] Nicolini G, Fogliata A, Vanetti E, Clivio A, Cozzi L. GLAaS: An absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications. Med Phys 2006;33(8):2839-2851.