ESTRO 2024 - Abstract Book
S3307
Physics - Detectors, dose measurement and phantoms
ESTRO 2024
Table 1. Penumbra widths of 5 cm x 5 cm profiles, calculated on both sides (left (L) and right (R)), before- and after replacing the x-ray target.
Penumbra X-direction
Penumbra Y-direction
Left [mm]
Right [mm]
Left [mm]
Right [mm]
Before
After
Before
After
Before
After
Before
After
3.34
3.23
3.39
3.30
3.95
4.03
4.03
4.11
Ratio After/Before
Ratio After/Before
Ratio After/Before
Ratio After/Before
0.968
0.972
1.019
1.020
Conclusion:
This work has shown that 6 MV small-field lateral dose profiles are impacted in the flattened region when the x-ray target is replaced. The penumbra of larger fields indicates a decrease in spot size in the X-direction and an increase in spot size in the Y-direction. OF were unchanged before and after x-ray target replacement and this was attributed to the effect of the increase/decrease of the spot size, cancelling each other out. The DLG was shown to increase after the replacement and this was attributed to the decrease of spot size in the X-direction. It is recommended to perform a rigorous verification after replacing the x-ray target, especially for linacs dedicated for small-field treatments.