Figure 1: Differences between DVH parameters for a prostate case, calculated by AAA, AXB w/m , RS, DC, and the corresponding MC w/m .

Conclusion:

Quantifying differences between 6FFF clinical dose distributions and corresponding distributions obtained by the software for pre-treatment dose control may assist identification of inaccuracies in the dose determination. The work may help establish action levels in the clinical workflow.

Keywords: Monte Carlo, Pre-treatment QC

References:

[1] Report of AAPM Task Group 219 on independent calculation- based dose/MU verification for IMRT. Med Phys 2021;48:e808-e829. DOI: 10.1002/mp.15069

[2] R. Chakarova, R. Cronholm, M. Krantz, P. Andersson, A. Hallqvist. An automated Monte Carlo QC system for volumetric modulated arc therapy: possibilities and challenges. Physica Medica 51 (2018) 32-37

1205

Digital Poster

Comparing Monte Carlo and Collapsed Cone algorithms for ArcCHECK analysis of complex treatment plans