Physics - Dose calculation algorithms

ESTRO 2024

Conclusion:

From the results for simple rectangular fields, we conclude that a homogeneous rather than heterogeneous phantom image appears to be more suited for ArcCHECK verification with MC. For these simple fields, no performance benefit from either MC or CC is found. However, MC might offer performance benefits over CC for more complex treatment plans. Results from clinical treatment plans suggest that our MC model performs comparable to our CC model for 6 MV. Notably, for 6 MV FFF, MC resulted in consistently better ArcCHECK results (10% to 58% lower mean gamma) as compared to CC.

Keywords: Monte Carlo algorithm, ArcCHECK, RayStation

1228

Digital Poster

Independent dose recalculation-based patient specific QA using ProBeam proton treatment log files.