S3430
Physics - Dose calculation algorithms
ESTRO 2024
Conclusion:
In phantom, AAA dose calculation accuracy in GGO-lesions is significantly correlated with the CTR. The larger the GGO component, the worse the performance of the algorithm. In patient, however, the edge between the solid and GGO component is less defined, and AAA calculation accuracy does not seem to significantly correlate with the CTR. Thus, in clinical practice, no additional precaution is needed for the dose calculation of non-solid pulmonary nodules.
Keywords: SBRT, Lung, Ground-Glass-Opacities
1785
Digital Poster
Comparison between Ethos measurements and dosimetric parametres calculated with Acuros XB Beam Model
Francisco Camba, Sergio Honorato, Nuria Perez-Gomez, Ismael Sancho-Kolster, Roberto Martin-Vaello, Manuel de la Cruz, Alex Pages-Perez, Cristina Picon
Catalan Institute of Oncology (ICO), Medical Physics and Radiation Protection, Barcelona, Spain
Purpose/Objective:
One of the most critical variables in ensuring the correct delivery of external radiotherapy treatment is the agreement between the dosimetric parameters measured in the treatment unit and those calculated by the Treatment Planning