ESTRO 2024 - Abstract Book
S3446
Physics - Dose calculation algorithms
ESTRO 2024
Variant H1 resulted in the irradiation of almost all the circulating blood (V >0Gy = 99.8 ± 0.7%) at the end of VMAT treatment, at a mean dose of 309.9 ± 74.7 mGy (Table 1 and Fig. 1). Taking into account lymphocytes recirculation in H2 variant, the fraction of lymphocytes receiving a non-zero dose dropped to 40.4 ± 10.2%, with a lower mean dose estimated at 52.6 ± 21.1 mGy, especially because irradiated lymphocytes passed through the irradiation field only 1.58 ± 0.91 times on average due to slow recirculation between lymphoid organs in the M 1 model. The most realistic scenario, including the OOF dose component (variant H3), showed that almost all the lymphocytes (97.6 ± 2.5%) were irradiated at the end of the VMAT treatment, mostly due to the OOF dose (mean dose equal to 265.6 ± 48.5).
Average dose
D2%
D50%
V >0.125Gy (%)
Model V >0Gy (%)
(mGy)
(mGy)
(mGy)
to all lymphocytes
to all lymphocytes
to all lymphocytes
to V>0Gy
to V>0Gy
to V>0Gy
87.6 ± 10.4
699.1
±
H1
99.8 + 0.7
309.9 ± 74.7 309.9 ± 74.7
285.4 ± 71.3 285.9 ± 71.1
699.4 ± 215.3
215.0
3.5 ± 2.1
H2
40.4 + 10.2
19.8 ± 4.8
52.6 ± 21.1
0.4 ± 0.9
34.3 ± 15.6
161.8 ± 55.6 223.7 ± 116.5
Made with FlippingBook - Online Brochure Maker