Conclusion:

Given the high dependence on the MLC- or collimator-defined aperture sizes, due acknowledgment should be made to accurate beam modelling alongside adequate linac commissioning and ongoing quality assurance (4) in this demanding setting. However, this novel work demonstrates that even for the smallest and most complex SRS targets it is possible to achieve highly accurate and clinically acceptable delivered dose gradients using a standard Truebeam linear accelerator without the use of a high-definition MLC using treatment plans produced in Varian Eclipse without the need for a separate beam model.

Keywords: SRS HyperArc complex

References:

1 Barbour AB, Jacobs CD, Williamson H, Floyd SR, Suneja G, Torok JA, Kirkpatrick JP. Radiation Therapy Practice Patterns for Brain Metastases in the United States in the Stereotactic Radiosurgery Era. Adv Radiat Oncol. 2019 Jul 26;5(1):43-52. doi: 10.1016/j.adro.2019.07.012. 2 David J. Eaton, Jonathan Lee, Ian Paddick,Stereotactic radiosurgery for multiple brain metastases: Results of multicenter benchmark planning studies. Pract Radiat Oncol. 2018; 8(4):e212-e220. https://doi.org/10.1016/j.prro.2017.12.011. 3 Paddick I, Mott J, Bedford J, Filatov P, Grishchuk D, Orchin G, Houston P, Eaton DJ. Benchmarking Tests of Contemporary SRS Platforms: Have Technological Developments Resulted in Improved Treatment Plan Quality? Pract Radiat Oncol. 2023 Sep-Oct;13(5):e451-e459. doi: 10.1016/j.prro.2023.05.005.