ESTRO 2024 - Abstract Book
S3722
Physics - Dose prediction, optimisation and applications of photon and electron planning
ESTRO 2024
The comparison map generated using Accumulation: ACTIVE Dosimetry was visually assessed to identify hot or cold areas of significance.
Results:
The V99% and V100% for CTVp_6500 can be found in the table below:
Patient
V99% planned
V99% delivered
V100% planned
V100% delivered
1
89.0
96.1
64.8
79.0
2
87.5
73.8
61.3
34.5
3
92.8
98.5
62.4
70.3
4
75.7
67.5
56.7
43.4
5
79.9
89.4
43.6
45.5
6
86.2
73.5
52.9
29.0
7
92.0
82.2
66.1
42.9
Additionally, all D95% doses delivered during treatment met the 61.3Gy minimum target, and the CTVp_6500 volumes receiving 95% of the prescription dose was 100% for 5/7 patients, with 99.6% and 99.9% for the other two patients.
Figure 2 below shows graphically the V100% for CTVp_6500 categorised by mouthbite use (y) or lack thereof (n).
Paired T tests were calculated for the D95%, V99% and V100% to compare planned dose verses delivered dose for all patients, additionally comparing patients treated with a mouthbite verses without. No tests were found to be statistically significant to p<0.05.