Conclusion:
Surface-guided setup with ETD is a viable alternative to traditional skin markers, and prepositioning using ETD SGRT has become a standard practice in our clinical workflow for all breast cancer patients treated in free breathing. Furthermore, we were able to use surface data to track the intrafraction motion with respiration excluded. Systematic movement was observed in the vertical direction, but it was considered negligible in a clinical setting.
Keywords: SGRT, Breast Cancer, Intrafraction Motion
840
Digital Poster
Experimental 4D validation of clinical control infrastructure for thorax patients in proton therapy.
Pietro Pisciotta, Sabine Visser, Erik W. Korevaar, Johannes A. Langendijk, Stefan Both
University Medical Center Groningen, Department of Radiotherapy, Groningen, Netherlands
Purpose/Objective:
A major challenge for high-precision pencil beam scanning (PBS) treatment in proton therapy (PT) is the respiratory induced target and organ motion. To treat moving targets with PBS-PT, it is mandatory to implement dedicated