ESTRO 2025 - Abstract Book
S2584
Physics - Detectors, dose measurement and phantoms
ESTRO 2025
Figure. The profile function (a) for electron beams with a nominal energy of 4 MeV formed using applicators with 4, 5, 6 and 10 cm diameters, and the percentage depth doses (PDD) (b) for electron beams with nominal energies of 4, 6, 9 and 12 MeV, formed using applicators with a diameter of 5 cm. Conclusion: The dosimetric measurements confirm the AQURE accelerator meets the accuracy standards outlined in the literature. References: [1] Beddar AS, et al. Intraoperative radiation therapy using mobile electron linear accelerators (TG-72). Med Phys 2006; 33(5):1476-89. doi:10.1118/1.2194447 [2] Istituto Superiori di Sanità. Guidelines for quality assurance in intra-operative radiation therapy. Oncología (Barc.) 2004; 27(7); doi:10.4321/S0378-48352004000700013 [3] Kruszyna-Mochalska M, et al. Zalecenia Polskiego Towarzystwa Fizyki Medycznej dotyczące kontroli jakości w radioterapii śródoperacyjnej promieniowaniem elektronowym (IOERT) za pomocą mobilnych akceleratorów. Inż i Fiz Med 2019; 8(1):7-25. [4] Rozp. Ministra Zdrowia z dnia 12 grudnia 2022 r. w sprawie testów eksploatacyjnych urządzeń radiologicznych i urządzeń pomocniczych. Dz.U. RP, 27 grudnia 2022, Poz. 2759. Keywords: IORT, dosimetry, commissioning