Conclusion: As shown through simulations, the detector demonstrated satisfactory performance for applications in FLASH 3D dosimetry, with a mean discrepancy below 0.01 a.u. and the largest deviations observed at the edges. Although some artifacts appeared in the reconstructed image due to the limited number of projections, the overall results are promising. Further experimental tests are planned in the coming months for comprehensive characterization and validation of the reconstruction algorithm using experimental data.

Keywords: FLASH radiotherapy, plastic scintillator dosimeter

3415

Digital Poster Characterization of Small Fields Using Scintillation Detectors Pedro Matias Liñán Rodríguez 1 , Victor González González 2 , Patricia Sánchez Rubio 1 , Luis Alejo Luque 3 , Aida López Romero 4 , Jose Luis Colado Tornero 1 , Ruth Rodríguez Romero 1 , Jaime Martínez Ortega 1 1 Medical Physics, Hospital Universitario Puerta de Hierro, Majadahonda, Spain. 2 Medical Physics, Instituto Valenciano de Oncología, Valencia, Spain. 3 Medical Physics, Hospital Gregorio Marañón, Madrid, Spain. 4 Medical Physics, Hospital Universitario de Guadalajara, Guadalajara, Spain