Physics - Image acquisition and processing
ESTRO 2025
Conclusion: The reconstructed eFOV impacted on SPR within 2.4%. The eFOV CT images were reliable for proton dose planning when the treatment area was positioned at the center of the scan field.
Keywords: extended FOV, proton therapy, stopping power ratio
References: 1. Wu RY, Williamson TD, Sahoo N, Nguyen T, Ikner SM, Liu AY, et al. Evaluation of the high definition field of view option of a large-bore computed tomography scanner for radiation therapy simulation. Phys Imaging Radiat Oncol. 2020;13:44-9. doi: 10.1016/j.phro.2020.03.004. 2. Lin CS, Tsai YC, Chen LH, Wang CW, Wu CJ, Chen WY, et al. Effect of extended field-of-view approaches on the accuracy of stopping power ratio estimation for single-energy computed tomography simulators. J Appl Clin Med Phys. 2023;24:e14010. doi: 10.1002/acm2.14010 3. Ainsley CG, Yeager CM. Practical considerations in the calibration of CT scanners for proton therapy. J Appl Clin Med Phys. 2014;15:4721. doi: 10.1120/jacmp.v15i3.4721.