ESTRO 2025 - Abstract Book
S3618
Physics - Quality assurance and auditing
ESTRO 2025
Figure 1: Histogram for both left and right banks in Elekta Versa HD nº2
The displayed graphs indicate that the majority of leaf deviations are below 0.15 mm, remaining well within tolerance limits and showing a higher concentration at smaller deviations. As suggested by literature of statistical process control for radiotherapy quality assurance [2], behavior charts were applied to each nominal position of the picket fence and globally for all positions. Figure 2 displays mean values for all nominal positions.
Figure 2: Behavior charts for mean values of all positions of the picket fence in Elekta Versa HD nº2
Previous graphics reflect the stability of MLC Agility, with the vast majority of points falling within control limits, and corrective actions taken when predefined values were exceeded.
Control limit
Lower left
Upper left
Lower right
Upper right
VERSA 1
-0.065
0.269
-0.136
0.233
VERSA 2
-0.196
0.287
-0.240
0.253
VERSA 3
-0.158
0.261
-0.057
0.315
Data table: Control limits for all three Elekta Versa HD
Control limits were found to be lower than the tolerance; however, in some cases, only small differences between the three machines were observed.
Conclusion: The Elekta Agility MLC model demonstrates excellent performance, as indicated by the tight control limits, which are