Physics - Quality assurance and auditing
ESTRO 2025
Conclusion: We conducted a comprehensive patient-specific QA for LRT treatment delivery dose verification through 3D-array diode measurements, TLF-based delivered dose calculation, and independent IROC TLD and film evaluations. Following TG218 guidelines with γ (3%,2mm) criteria and tolerance/action limits of 95%/90%, most ArcCHECK/Delta4 measurements exceeded the 95% tolerance, with all measurements meeting the action threshold.
Keywords: 3D-array phantom, Monte-Carlo Dose calculation
References: [1] Heng VJ, (2024) Towards a clinical implementation of mixed electron-photon radiation therapy. [Doctoral thesis, McGill University].
1684
Digital Poster A new derivation of the gamma index for signed data analysis and advanced quality assurance in radiotherapy JImmy Fontaine 1 , Abdourahamane Djibo-Sidikou 2 , Maud Suszko 1 , Florian Mahinc 1 , Agathe Leroux 1 , Clemence Monnet 1 , Julien Langrand-Escure 3 , Bertrand Fleury 3 , Paul Retif 2,4