ESTRO 2025 - Abstract Book
S3636
Physics - Quality assurance and auditing
ESTRO 2025
Results:
Figure 1. Jaw defined small-field OF comparison
The OF comparisons show: 1) the EPID-derived and measured OFs are in close agreement apart from at the 1x1 cm 2 field where EPID is over-estimating; 2) OFs for both measured and EPID show small variation between the centres for above 0.5x0.5 cm 2 field size with ~1.5% (1 SD) variation for jaw and ~1.0% for MLC at 1x1 cm 2 field size ; 3) variations in the TPS beam-models between the centres are apparent for both jaw and MLC defined field, with models over-predicting the smallest field OF for jaw and under-predicting the smallest OF for MLC. TPS calculated OFs at 1x1 cm 2 vary more than measured OFs between the centres.
Figure 2. MLC defined small-field OF comparison
Conclusion: These data suggest that an EPID method can reproduce measured doses for very small fields used in MTSI SRS without the need for specialised equipment, resource-intensive measurement techniques or an onsite auditor. These data will also enable standardisation between institutions to consolidate both MLC setup and TPS dose algorithms for MTSI SRS.
Made with FlippingBook Ebook Creator