Actions included the implementation of proactive DLP assessment prior to CT scanning for all patients, altering scan parameters if possible to reduce DLP to within 10% of reference. Consultants must also specify exact parameters when protocol exceeds routine scan parameters. Conclusion: Annual review of DRLs and application of continuous quality improvement plans ensure good practice in the absence of national DRLs to audit against. We currently have four years’ CT data in our newly established centre. Annual review of radiotherapy CT protocols and the cycle of annual quality improvement has enabled us to optimise CT dose, promoting patient safety. References: 1. Health Information and Quality Authority. Diagnostic Reference Levels. Guidance on the establishment, use and review of diagnostic reference levels for medical exposure to ionising radiation, July 2021. [Online]. Available: https://www.hiqa.ie/reports-and-publications/guide/guidance-establishment-use-and-review-diagnostic-reference levels. 2. European Union Regulations 2018, SI No. 256 of 2019, 2018. [Online]. Available: http://www.irishstatutebook.ie/2018/en/si/0256.html. Keywords: Quality Improvement, Computed Tomography, DRLs