ESTRO 36 Abstract Book
S261 ESTRO 36 _______________________________________________________________________________________________
contour OAR located within 3cm from PTV opt
, based on the
daily MR imaging (SMART 3CM ). Optimization structures are automatically adapted to the new anatomy, and re- optimization is performed using exactly the same plan parameters. This limited re-contouring strategy was evaluated by comparing 45 previously delivered fractions against a simulated standard (re-)planning method using full-scale OAR (re-)contouring, where optimization objectives were used for the whole organs (SMART FULLOAR ). Baseline plans for OAR were created that had identical plan quality as were achieved for SMART 3CM . Efficiency of both strategies was scored according to the number of optimizations needed to generate a high quality plan. Plan quality was assessed using PTV coverage (V95%) and institutional OAR constraints (V33Gy and V25Gy).
Conclusion This fast and robust (re-)planning approach for SBRT to pancreatic tumors requires clinicians to only re-contour OARs located within 3cm of the PTV OPT. Spatially partitioned optimization structures within this 3 cm region allowed for optimal OAR sparing, and adequate target coverage, using exactly the same plan parameters. OC-0491 Quality assurance of a novel table mounted imaging device integrated in a patient positioning system A. Utz 1 , A. Ableitinger 1 , A. Zechner 1 , M. Mumot 1 , M. Teichmeister 1 , P. Steininger 2 , H. Deutschmann 2 , M. Stock 1 1 EBG MedAustron GmbH, Medical Physics, Wiener Neustadt, Austria 2 medPhoton GmbH, Medical Physics, Salzburg, Austria Purpose or Objective Image guided radiation therapy (IGRT) aims to reduce margins and subsequently increase dose sparing for OAR. The majority of image guidance procedures are based on ceiling/floor- or gantry mounted imaging devices. In our particle therapy center a novel approach for patient alignment was introduced. The imaging system (imaging ring) is mounted on the treatment table and as such, allows high imaging flexibility e.g. CBCT or planar imaging at different table positions. The goal was to establish a phantom and a concept for a quality assurance procedure for the whole IGRT workflow. Material and Methods The IGRT Phantom consists of a PMMA cube with steel fiducials, which can be placed in predefined offset positions on a baseplate to simulate clinical patient shifts. An additional support structure is used to lift the cube. Holes on the upper corners of the cube allow to independently determine the absolute position with a lasertracker (see figure 1a). A CT imageset of the cube in a reference position serves as planning CT. The baseplate is indexed on the patient couch. The cube can be moved to a predefined translational and rotational offset position on the couch. Two planar images were acquired and registered to the CT.The calculated correction vector was applied by the patient positioning system. This workflow was repeated at three index positions (equal distributed along the treatment volume), payloads (0kg, 100kg, 150kg), cube offsets (long.: 10mm, lat.: 15mm, vert.: -
Results The SMART 3CM
baseline plans required a lower number of
optimizations than SMART FULLOAR
(4 vs 17 on average).
PTV OPT coverage with both strategies was identical in all uniformly resulted in plans which complied with the V33Gy dose constraint for OARs, whereas SMART FULLOAR failed in 35% of the cases to adhere to the V33Gy dose constraint according to the clinical protocol. Both strategies achieved V25Gy values lower than 20 cc for all OARs in every fraction. However, on average, SMART 3CM resulted in a lower V25Gy than SMART FULLOAR (Fig.2). fractions (median V95%=93±6.8%). SMART 3CM
Made with FlippingBook